Sanjaa-Uyen-Tumur ophthalmometer: An introduction of the objective ophthalmometer


 Ophthalmology Department, Orbita Eye Hospital, Ulaanbaatar, Mongolia

Correspondence Address:
Uyen Bold,
Orbita Eye Hospital, Ulaanbaatar
Mongolia
Login to access the Email id

Source of Support: None, Conflict of Interest: None

DOI: 10.4103/tjo.tjo_58_19

PURPOSE: The purpose is to present the objective of the Sanjaa-Uyen-Tumur (SUT) ophthalmometer and compare its measurements with the standard Hertel exophthalmometer. METHODS: Eyeball position of 88 healthy eyes of the patients was measured using both the SUT ophthalmometer and Hertel exophthalmometer. Both methods were performed in one session by the same experienced ophthalmologist. Data were analyzed using Bland–Altman method. RESULTS: Mean age of the participants was 41.11 ± 14.08, further, 59% (26) were male and 41% (18) were female, respectively. The mean difference was 0.1420455 and the standard deviation of the difference was 0.9221067, 95% confidence interval, respectively. Based on the Bland–Altman analysis, lower limit of agreement in our study was from −2.284538 to −1.614211 and upper limit of agreement was from 1.33012 to 2.000447. According to the results of our studies, there is no relationship between the difference and the level of measurement in either plot. CONCLUSION: This SUT ophthalmometer can be used same as the Hertel exophthalmometer in ophthalmology practice as can be accurate, affordable, and objective.


Print this article
Search
 Back
 
  Search Pubmed for
 
    -  Bold U
    -  Tumur S
    -  Kh O
    -  Sanjaa E
 Citation Manager
 Article Access Statistics
 Reader Comments
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed159    
    PDF Downloaded4    

Recommend this journal