ORIGINAL ARTICLE
Year : 2017  |  Volume : 7  |  Issue : 3  |  Page : 149-154

Intraocular pressure monitoring by rebound tonometry in children with myopia


1 Department of Ophthalmology, Taipei City Hospital, Taipei, Taiwan
2 Department of Ophthalmology, Taipei City Hospital; Community Medicine Research Center, Institute of Public Health, National Yang-Ming University, Taipei, Taiwan
3 Department of Ophthalmology, Taipei City Hospital; Department of Ophthalmology, National Taiwan University Hospital, Taipei, Taiwan
4 Department of Ophthalmology, Taipei City Hospital; Department of Ophthalmology, College of Medicine, National Yang-Ming University, Taipei, Taiwan

Correspondence Address:
Ya-Chuan Hsiao
Department of Ophthalmology, Taipei City Hospital; College of Medicine, National Yang-Ming University, Taipei
Taiwan
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/tjo.tjo_45_17

Rights and Permissions

Background/Purpose: Topical atropine treatment is generally accepted to retard the progression of myopia, but it is associated with side effects such as photophobia and elevation of intraocular pressure (IOP). IOP measurements in children are challenging. The traditional applanation tonometry by direct contact with the cornea will require patient's cooperation. The rebound tonometer, using a dynamic electromechanical method for measuring IOP, shows good correlation with traditional tonometry. The purpose of this study is to evaluate the IOP of myopic children under atropine treatment using rebound tonometer and to compare the characteristics between rebound tonometry and applanation tonometry. Methods: This study is a prospective study measuring IOP by rebound tonometer in myopic children under regular low-dose atropine treatment. We recruited children with refraction error showing myopia over −0.5 D with 0.15%, 0.3%, or 0.5% atropine eye drops use every night or every other night for myopia control. Children with treatment duration of atropine <1 month were excluded from the study. IOP measurements were performed by applanation tonometer (Tono-Pen XL, Reichert) and rebound tonometer (ICARE). The reliability of rebound tonometer was analyzed with percentage. Comparison of IOP between rebound tonometer and applanation tonometry was presented. Results: The rebound tonometry was well tolerated by all participants and caused no complaints, discomfort, or adverse events. Totally 42 myopic eyes of 42 subjects were included in the study. The average age of these participants was 10 years old, range from 5 to 16. Median = 10 years old. The average IOP of the right eye by rebound tonometer was 17.4 ± 3 mmHg, and 17.1 ± 3 mmHg by applanation tonometry. Nearly 19%, 33%, and 24% of difference of IOP readings between rebound tonometer and Tono-Pen applanation are within 0 mmHg, 1 mmHg, and 1–2 mmHg, respectively. Conclusions: Rebound tonometry has good correlation with applanation tonometry and 76.1% of differences between two tonometers are <2 mmHg. The advantage of drop-free rebound tonometry has made it easier to obtain IOP readings in myopia children under atropine treatment.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed3760    
    Printed123    
    Emailed0    
    PDF Downloaded391    
    Comments [Add]    
    Cited by others 3    

Recommend this journal