CASE REPORT
Year : 2018  |  Volume : 8  |  Issue : 1  |  Page : 49-51

Postoperative intraocular lens opacification


1 Department of Ophthalmology, National Taiwan University Hospital, University College of Medicine, Taipei, Taiwan
2 Department of Ophthalmology, Cathay General Hospital, Taipei, Taiwan

Date of Submission28-Mar-2017
Date of Acceptance14-Jul-2017
Date of Web Publication26-Mar-2018

Correspondence Address:
Yu-Chih Hou
Department of Ophthalmology, Cathay General Hospital, Taipei, Taiwan, 280, Ren-Ai Road Sec 4, Taipei
Taiwan
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/tjo.tjo_78_17

Rights and Permissions
  Abstract 


Intraocular lens (IOL) opacification is rare but may occur after non-Descemet stripping automated endothelial keratoplasty (n-DSAEK) or intravitreal air injection after pars plana vitrectomy (PPV). We reported two cases of IOL opacification within the pupillary region. Chart was retrospectively reviewed. The predisposing factors and the visual acuity were analyzed. The opacification was evaluated by anterior segment optical coherence tomography (AS-OCT). A 68-year-old healthy woman with pseudophakic bullous keratopathy underwent uneventful n-DSAEK in the right eye. Postoperative vision was 20/40. Nine months after surgery, fine granular deposits were seen in the anterior surface of IOL. The vision decreased to 20/50 but remained stable during 3-year follow-up. A 61-year-old man with diabetes mellitus received PPV and silicone oil tamponade for retinal detachment and vitreal hemorrhage after cataract surgery in the right eye. Removal of silicone oil and intravitreal air injection was performed, and postoperative vision was 20/100. Granular deposits were observed in hydrophobic acrylic IOL 1 month after surgery. The visual acuity decreased to 20/120. AS-OCT revealed hyperreflective materials in the anterior surface of IOL in both cases. An uncommon phenomenon of IOL opacification in the pupil region may occur after n-DSAEK or PPV, which may be associated with intraocular air injection or systemic diseases.

Keywords: Intraocular air injection, intraocular lens, non-Descemet stripping automated endothelial keratoplasty, opacification


How to cite this article:
Ma ST, Yang CM, Hou YC. Postoperative intraocular lens opacification. Taiwan J Ophthalmol 2018;8:49-51

How to cite this URL:
Ma ST, Yang CM, Hou YC. Postoperative intraocular lens opacification. Taiwan J Ophthalmol [serial online] 2018 [cited 2023 Mar 27];8:49-51. Available from: https://www.e-tjo.org/text.asp?2018/8/1/49/217810




  Introduction Top


Intraocular lens (IOL) opacification has been observed with most materials.[1] Its presentation may be various and the causes may be multifactorial. Descemet stripping automated endothelial keratoplasty (DSAEK) provides many advantages over traditional penetrating keratoplasty. However, an uncommon complication of IOL opacification after DSAEK was reported recently and could impair vision.[2],[3],[4],[5] This phenomenon may also occur in nonkeratoplasty cases such as filtration surgery for glaucoma and pars plana vitrectomy (PPV).[6],[7] Here, we reported two cases of opacification in the anterior surface of IOL limited to the pupillary area. One case underwent non-DSAEK (n-DSAEK), and the other case received removal of silicone oil and intravitreal air injection.


  Case Reports Top


Case 1

A 68-year-old woman was well healthy and underwent phacoemulsification and IOL implantation in the right eye at a local clinic in March 2012. Postoperative vision decreased, and she visited to our clinic 1 month later. Biomicroscopy revealed corneal edema with a vision of 20/200 in the right eye. Then, an uneventful n-DSAEK was smoothly performed in her right eye 4 months later. A donor corneoscleral button was mounted on an artificial anterior chamber and dissected with automated microkeratome (Moria, Inc, France). After 8-mm trephination, a thin lenticule was placed on Busin glide and pulled into anterior chamber by forceps, where it was unfolded and attached to the recipient cornea by air bubble for 10 min. The final air was left as the graft size at the end of surgery. Postoperatively, the lenticule was well attached to the cornea. Topical 0.1% betamethasone was prescribed four times daily and was tapered to twice daily 2 months later. Postoperative vision improved to 20/40. Nine months after surgery, a 3-mm circumscribed opacification was seen in the anterior surface of the IOL within the pupil region [Figure 1]a. The opacification was characterized by fine and white granular deposits. The vision of the right eye slightly decreased to 20/50. However, the opacification did not increase and her vision was 20/50 during the 3-year follow-up.
Figure 1: External eye photography. Fine granular deposits occurred in the anterior surface of the intraocular lens within the pupillary region 9 months after non-Descemet stripping automated endothelial keratoplasty in case 1 (a) and 1 month after intravitreal air injection in case 2 (b)

Click here to view


Case 2

A 61-year-old man with diabetes mellitus underwent uneventful phacoemulsification and implantation of IOL (AcrySof SA60AT; Alcon Surgery) in the right eye in December 2013. Later, he presented with proliferative diabetic retinopathy, retinal detachment, and vitreous hemorrhage with a vision of hand motion in the right eye. PPV and silicone oil tamponade were performed in November 2014. Postoperative course was smooth and vision was 20/120. Removal of silicone oil and intravitreal air injection was performed in March 2016, and postoperative vision was 20/100. However, whitish granular deposits were observed in the anterior surface of the IOL confined to the pupillary zone 1 month after surgery and vision decreased to 20/120 [Figure 1]b. The opacification and his vision were stationary in the 9-month follow-up.

Anterior segment optical coherence tomography (AS-OCT; Cirrus HD-OCT; Carl Zeiss Meditec Co. Ltd.; Tokyo, Japan) revealed hyperreflective opacity in the anterior surface of the IOLs in the two cases [Figure 2]a and [Figure 2]b.
Figure 2: Anterior segment optical coherence tomography. In spite of the hyperreflective signal can be observed between different media in anterior segment optical coherence tomography image, the signal is more intense in the anterior surface of the Intraocular lens in the pupillary area, which indicated the area of material deposition in case 1(a) and case 2 (b)

Click here to view



  Discussion Top


The opacification of the IOLs has been observed with some specific materials, or manufacturing or packaging process, including a snowflake opacification in polymethylmethacrylate IOLs, discoloration in silico ne IOLs, calcification in hydrophilic acrylic IOLs, and microvacuoles (glistenings) in hydrophobic IOLs.[8],[9] IOL material and structure may have an impact on the affinity of calcification. Most severe cases of IOL opacification were associated with hydrophilic or hydrophilic-hydrophobic IOLs, while it could also be observed in hydrophobic IOL though relatively rare.[10] The opacification may present in both anterior and posterior surface, deep central region, haptics, or nearly whole lens.[1],[6],[11]

Since endothelial keratoplasty became popular surgery for corneal endothelial dysfunction, a rare complication of IOL opacificaton was observed. The opacification showed a particular presentation of numerous fine granular deposits in the anterior surface of IOL limited to the pupil zone, which could be demonstrated by Scheimpflug photography [4],[7] or AS-OCT. The major reported DSAEK cases were hydrophilic acrylic IOL, especially in those who experienced rebubbling.[2],[3],[4],[5] However, some of nonkeratoplasty cases with intravitreal injection of air or gas after PPV were also reported to have a similar pattern of IOL opacification.[6],[7] Case 2 showed the opacification could also occur in hydrophobic IOL, though his presentation was not the same as the previously reported cases. The duration between the onset of IOL calcification and surgery in most reported cases ranged from 1 month to 6 years.[2],[3],[4],[5],[6],[7],[8],[9],[10],[11] The deposits were mainly composed of calcium and phosphate, which were confirmed by special stains, scanning electron microscopy, and energy-dispersive x-ray spectroscopy.[2],[3],[4],[5],[6],[7]

The exact mechanism for IOL opacification still remains speculative and is believed to be multifactorial. Numerous potential risk factors have been proposed, such as hydrophilic acrylic IOL exposure to air, patient's general medical condition, or breakdown of the blood-aqueous barrier (BAB) during intraocular surgery, excessive postoperative ocular inflammation.[2],[3],[4],[5] Patients' systemic factors included renal failure and diabetes mellitus.[10],[11] The presence of intracameral air or gas contact with IOL surface may enhance subsequent crystallization under high concentration of calcium in the anterior chamber. Prolonged breakdown of the BAB may result in metabolic change of the microenvironment of the anterior chamber, including increase in protein, cells, and calcium content. Because the calcium deposits were inside the surface of the IOL and was hard to be removed by irrigation or polish, some cases may need IOL exchange.[2],[3],[4],[5],[6],[7]

Case 1 was only one case among 183 DSAEK or n-DSAEK cases in National Taiwan University Hospital from July 2009 to December 2016. There was no any known systemic factor or surgical complication contributed to the possible metabolic alteration in the aqueous in case 1. Her cataract surgery was performed elsewhere, so the implanted IOL material was unknown. Case 2 with diabetes mellitus had IOL opacification 1 month after intravitreal air injection. Both cases shared a similar situation of intraocular air injection, which might cause electrolytes imbalance and calcium deposits in the anterior surface of IOL. Other related factors, such as IOL materials, systemic diseases, or postoperative chronic inflammation, might also play a role in the development of their IOL calcification, especially in case 2. Because the vision impairment was relatively mild and remained stable in these 2 cases, their IOLs were not explanted. We could not confirm the true cause and material deposits in the opacification of the 2 patients, but calcium deposits were the most probable according to literature review. Surgeons should be aware of this uncommon phenomenon of IOL calcification and thoroughly investigated this condition by taking detailed medical and surgical history, the extent of vision impairment, the examination of AS-OCT and the serial records of IOL opacity, which could aid in avoidance of mismanagement such as the attempt of opacity removal by neodymium: Yttrium-aluminum-garnet laser or invalid surface polish of IOL by irrigation and aspiration. However, some severe cases may need IOL exchange.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form, the patient has given their consent for their images and other clinical information to be reported in the journal. The patient understands that name and initials will not be published and due efforts will be made to conceal identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

The authors declare that there are no conflicts of interests of this paper.



 
  References Top

1.
Neuhann IM, Werner L, Izak AM, Pandey SK, Kleinmann G, Mamalis N, et al. Late postoperative opacification of a hydrophilic acrylic (hydrogel) intraocular lens: A clinicopathological analysis of 106 explants. Ophthalmology 2004;111:2094-101.  Back to cited text no. 1
[PUBMED]    
2.
Dhital A, Spalton DJ, Goyal S, Werner L. Calcification in hydrophilic intraocular lenses associated with injection of intraocular gas. Am J Ophthalmol 2012;153:1154-60.e1.  Back to cited text no. 2
    
3.
Neuhann IM, Neuhann TF, Rohrbach JM. Intraocular lens calcification after keratoplasty. Cornea 2013;32:e6-10.  Back to cited text no. 3
[PUBMED]    
4.
Fellman MA, Werner L, Liu ET, Stallings S, Floyd AM, van der Meulen IJ, et al. Calcification of a hydrophilic acrylic intraocular lens after Descemet-stripping endothelial keratoplasty: Case report and laboratory analyses. J Cataract Refract Surg 2013;39:799-803.  Back to cited text no. 4
    
5.
De Cock R, Fajgenbaum MA. Calcification of Rayner hydrophilic acrylic intra-ocular lenses after Descemet's stripping automated endothelial keratoplasty. Eye (Lond)2014;28:1383-4.  Back to cited text no. 5
    
6.
Walker NJ, Saldanha MJ, Sharp JA, Porooshani H, McDonald BM, Ferguson DJ, et al. Calcification of hydrophilic acrylic intraocular lenses in combined phacovitrectomy surgery. J Cataract Refract Surg 2010;36:1427-31.  Back to cited text no. 6
    
7.
Cao D, Zhang H, Yang C, Zhang L. Akreos Adapt AO intraocular lens opacification after vitrectomy in a diabetic patient: A case report and review of the literature. BMC Ophthalmol 2016;16:82.  Back to cited text no. 7
    
8.
Apple DJ, Peng Q, Arthur SN, Werner L, Merritt JH, Vargas LG, et al. Snowflake degeneration of polymethyl methacrylate posterior chamber intraocular lens optic material: A newly described clinical condition caused by unexpected late opacification of polymethyl methacrylate. Ophthalmology 2002;109:1666-75.  Back to cited text no. 8
    
9.
Werner L. Glistenings and surface light scattering in intraocular lenses. J Cataract Refract Surg 2010;36:1398-420.  Back to cited text no. 9
    
10.
Prasad PS, Oliver SC, Gonzales CR. Midlenticular optic opacification of a hydrophobic acrylic intraocular lens in association with retained perfluorocarbon liquid following vitreoretinal surgery. Ophthalmic Surg Lasers Imaging 2010;9:1-2.  Back to cited text no. 10
    
11.
Pandey SK, Werner L, Apple DJ, Kaskaloglu M. Hydrophilic acrylic intraocular lens optic and haptics opacification in a diabetic patient: Bilateral case report and clinicopathologic correlation. Ophthalmology 2002;109:2042-51.  Back to cited text no. 11
    


    Figures

  [Figure 1], [Figure 2]


This article has been cited by
1 Opacification of refractive bifocal intraocular lens in one month
Yanfeng Zeng, Min Liang, Licheng Fan, Sen Xu, Fengting Liu, Xiaoli Zhou, Xin Tan, Xiaoqin Wang
Medicine. 2022; 101(5): e28757
[Pubmed] | [DOI]
2 A review of late intraocular lens opacifications
Piotr Kanclerz,Timur M. Yildirim,Ramin Khoramnia
Current Opinion in Ophthalmology. 2021; 32(1): 31
[Pubmed] | [DOI]
3 Clinical and material degradations of intraocular lenses: A review
Nick Stanojcic,Christopher Hull,David PS O’Brart
European Journal of Ophthalmology. 2020; 30(5): 823
[Pubmed] | [DOI]
4 Systematic review of potential causes of intraocular lens opacification
Joaquín Fernández,Alicia Sánchez-García,Manuel Rodríguez-Vallejo,David P. Piñero
Clinical & Experimental Ophthalmology. 2019;
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Case Reports
Discussion
References
Article Figures

 Article Access Statistics
    Viewed4888    
    Printed161    
    Emailed0    
    PDF Downloaded482    
    Comments [Add]    
    Cited by others 4    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]