ORIGINAL ARTICLE
Year : 2021  |  Volume : 11  |  Issue : 2  |  Page : 168-174

Investigation of the repeatability of tear osmolarity using an I-PEN osmolarity device


1 Department of Optometry, College of Applied Medical Sciences, King Saud University, Riyadh, Saudi Arabia
2 Cornea Research Chair, Department of Optometry, College of Applied Medical Sciences, King Saud University, Riyadh, Saudi Arabia

Correspondence Address:
Dr. Raied Fagehi
Department of Optometry, College of Applied Medical Sciences, King Saud University, P O Box 10219, Riyadh 11433
Saudi Arabia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/tjo.tjo_65_20

Rights and Permissions

PURPOSE: To investigate the repeatability of tear osmolarity in healthy Saudi subjects using an I-PEN osmolarity device. MATERIALS AND METHODS: Thirty typical male subjects with healthy eyes (27.4 ± 4.9 years) participated in the study. Eye abnormalities were tested with a slit lamp, and eye comfort was determined with the surface disease index. Measurements of the tear break-up time and phenol red thread tests were used for as exclusion criteria. The tear osmolarity test, using an I-PEN osmolarity system, was performed three times in the right eye of each subject with a 5 min' gap between tests. RESULTS: The average osmolarity test score was 303.8 ± 4.8 mOsm/L. Tear osmolarity measurements showed tear osmolarity of 280–299 mOsm/L, 300–309 mOsm/L, and 310–329 mOsm/L in 14 (46.7%), three (10%), and 13 (43.3%) subjects, respectively. Correlations among the three I-PEN measurements were significant (Spearman's correlation coefficient; r = 0.036, 0.501, and 0.603; P = 0.050, 0.006, and 0.001, respectively). The mean coefficient of variance among the three measurements was 4.4%. CONCLUSION: The mean measurement of an I-PEN tear osmolarity was 303.8 ± 4.8 mOsm/L which is in agreement with the range of those reported for healthy subjects. The I-PEN is reliable and has the advantage of portability (hand-held) compared to the other osmolarity systems.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1095    
    Printed14    
    Emailed0    
    PDF Downloaded59    
    Comments [Add]    

Recommend this journal